Method of Analysis at the U.S. Geological Survey California Water Science Center, Sacramento Laboratory – Determination of Haloacetic Acid Formation Potential, Method Validation, and Quality-Control Practices
Scientific Investigations Report 2005-5115, published 2005.
Figure 1
Chromatograms for (A) level-5 calibration standard, (B) extraction blank, and (C) full procedural blank.
Peaks are labeled with analyte name and retention time in minutes. IS, internal standard.
Chromatograms for (A) level-5 calibration standard, (B) extraction blank, and (C) full procedural blank.
 
This method is designed to be used on water samples to determine the HAAFP under controlled, standard conditions of pH, temperature, darkness, contact time between Cl2 and the water sample, and residual Cl2. Because the HAAFP depends on the experimental conditions, data generated from this method only should be compared to data generated under the same experimental conditions. This method was modified from the USEPA Method 552.2 to determine HAA in drinking water, surface water, or ground water (table 1), and is suitable for filtered natural waters and experimentally produced water samples. The calibration range for each of the nine HAA is listed in table 1. Water samples that produce higher concentrations of HAA should be diluted prior to analysis.
Water samples are collected and filtered to remove suspended particulate matter. The DOC and ammonia-nitrogen (NH3-N) concentrations in each water sample are used to determine the appropriate amount of Cl2 dose solution to add. The samples are dosed with sufficient Cl2 to satisfy sample Cl2 demand and leave a residual Cl2 concentration of 2–4 milligrams per liter (mg/L). A buffer is added to maintain a sample pH of 8.3 during chlorination. After the samples are dosed, they are incubated in the dark for 7 days at 25°C (77°F).
At the end of the 7 days, the pH and residual-free Cl2 are measured and the samples are quenched with sodium sulfite (Na2SO3) solution to neutralize any remaining free Cl2. A 40 milliliter (mL) volume of sample is adjusted to pH < 0.5 and extracted with 2 mL of methyl tert-butyl ether (MTBE). The organic phase containing the HAA is separated, and the HAA are converted into their methyl esters by the addition of acidic methanol with slight heating. A second extraction using dilute sodium hydroxide concentrates the methyl ester compounds in the organic phase. The target analytes are identified and measured by GC-ECD. Analytes are quantified by using a procedural standard calibration curve. The primary differences between this method and the USEPA Method 552.2 are that different compounds are used for the surrogate and internal standard, smaller and silanized reaction vials are used, the final extract volume is half as much, and the second extraction is done with NaOH rather than sodium bicarbonate.
The equipment and materials used for analysis of HAAFP are listed below. The organic carbon-free water is produced on-site with a recirculation Picotech water system (Hydro Service and Supplies, Inc.). Inlet water for the Picotech water system is deionized and produced on-site with a Culligan deionizing system (Culligan International Company). Scheduled routine maintenance and replacement of cartridges are done on both systems. The organic carbon-free water is tested frequently by analysis of DOC and trihalomethane formation potential (Bird and others, 2003; Crepeau and others, 2004).
Equipment and materials used for analysis of haloacetic acid formation potential. (Product and firm names are listed for documentation purposes only.)
Sample Containers
  • Baked amber glass bottles with Teflon-lined lids
Ammonia and chlorine measurements
  • Ammonia salicylate and cyanurate reagent powder pillows (Hach, Loveland, Colorado)
  • Hach N-diethyl-p-phenylenediamine free-chlorine reagent powder pillows, catalog number 14077-28 or dispenser catalog number 10445 (Hach, Loveland, Colorado)
  • 2-dram Opticlear vials, screw thread, catalog number 60910-2 (Kimble Glass, Inc.)
  • Pipettes, 1- and 5-milliliter adjustable Oxford Benchmate (Nichiryo Co., LTD) with disposable plastic tips (Labsource, Fisherbrand, or equivalent)
  • Spectrophotometer, Genesys20 (ThermoSpectronic)
Dilution
  • Bottle-top dispenser, adjustable from 10 to 109 milliliter (Fisher/Wheaton, Pittsburg, Pennsylvania)
  • Glass beakers and graduated cylinders 25–100 milliliter (Fisher Scientific, Pittsburg, Pennsylvania)
  • Organic-free water, produced onsite with Pico-pure water system (Hydro Service and Supplies, Inc.)
  • Pipettes, 1- and 5-milliliter adjustable Oxford Benchmate (Nichiryo Co., LTD) with disposable plastic tips (Labsource, Fisherbrand, or equivalent)
Dosing and Quenching
  • Analytical balance, accuracy of 0.050 gram ±0.0001 gram
  • Boric acid (Mallinckrodt analytical reagent grade or equivalent)
  • Dilute hydrochloric acid and dilute sodium hydroxide for pH adjustment (0.1 N, Fisher Scientific, Pittsburg, Pennsylvania)
  • pH buffer 7 and 10 (U.S. Geological Survey Ocala Water-Quality Laboratory, Ocala, Florida)
  • pH meter, Orion model 420A with Triode gel electrode (Orion Research Inc., Beverly, Massachusetts)
  • Sodium hydroxide pellets (American Chemical Society reagent grade, Aldrich Chemical, Milwaukee, Wisconsin)
  • Sodium hypochlorite 4–6 percent (Fisher purified grade, Fisher Scientific, Pittsburg, Pennsylvania)
  • Sodium sulfite, anhydrous (American Chemical Society reagent grade, Fisher Scientific, Pittsburg, Pennsylvania)
  • 40-milliliter vials, amber borosilicate, TraceClean (VWR Scientific, West Chester, Pennsylvania)
Extraction
  • Copper II sulfate pentahydrate (Certified American Chemical Society, Fisher Scientific, Pittsburg, Pennsylvania)
  • Graduated cylinders (50 milliliter)
  • methyl tert-butyl ether (J.T. Baker, Mallinckrodt Baker, Inc., Phillipsburg, New Jersey)
  • Sodium sulfate (Certified American Chemical Society, Fisher Scientific, Pittsburg, Pennsylvania)
  • Sulfuric acid (American Chemical Society reagent grade, Fisher Scientific, Pittsburg, Pennsylvania)
  • 60-milliliter vials
Methylation
  • Autosampler vials, 4-milliliter amber, silanized (National Scientific Company)
  • Autosampler Target DP vials, amber with silanized inserts (National Scientific Company)
  • Glass pipettes (2 milliliter)
  • Ice bath
  • Sodium hydroxide diluted to 0.1 Normal (Fisher Scientific, Pittsburg, Pennsylvania)
  • Volumetric glassware
  • Water bath (50ºC)
  • 10 percent (volume/volume) sulfuric acid in pesticide grade methanol (Fisher Scientific, Pittsburg, Pennsylvania)
Haloacetic acid analysis
  • Autosampler, Hewlett-Packard 6890 (Wilmington, Delaware)
  • Column, Rtx-5, 30 meters in length, 0.25-millimeters internal diameter with 0.25-microns film thickness and Integra-Guard (guard column) Catalog number 10223-127 (Restek, Bellefonte, Pennsylvania)
  • Gas chromatograph, Hewlett-Packard 5890 (Wilmington, Delaware)
  • Nitrogen, ultra-high purity (Praxair, Sacramento, California)
Standards
  • 2-bromopropionic acid 1 mg/mL in methyl tert-butyl ether (Supelco, Bellefonte, Pennsylvania)
  • 2-bromo-1-chloropropane 2 mg/mL in methanol (Supelco, Bellefonte, Pennsylvania)
  • Nine haloacetic acid mix in methyl tert-butyl ether (Supelco, Bellefonte, Pennsylvania)
  • Neat solutions of individual haloacetic acids (Sigma-Aldrich, St. Louis, Missouri)
  • Volumetric glassware (5 milliliter, Fisher Scientific, Pittsburg, Pennsylvania)
Glassware is washed with Liquinox soap and rinsed with copious amounts of organic carbon-free water. The glassware openings are covered with aluminum foil and the glassware is baked in a muffle furnace at 450ºC for 4 hours. The baked glassware is stored with the foil still on in closed drawers or cabinets until needed.
The sodium sulfate (Na2SO4) is baked in a muffle furnace at 400ºC for up to 4 hours to remove phthalates and other potentially interfering organic substances. The baked Na2SO4 then is stored in a clean, capped glass bottle.
The 10 percent sulfuric acid (H2SO4)/methanol solution that is used for methylation must be prepared in a hood and with the appropriate personal protective equipment worn by the laboratory staff. This solution is prepared by adding 5 mL of concentrated H2SO4 drop-wise to 20–30 mL of methanol in a 50-mL volumetric flask. The flask should be placed in an ice bath during addition of the H2SO4 because the reaction is strongly exothermic. Once the solution has cooled, methanol is added to the volumetric flask to give a final volume of 50 mL.
A surrogate compound, 2-bromopropionic acid (CH3CHBrCO2H), is added to the samples to monitor the efficiency of extraction and methylation of HAA. This compound is chemically similar to the HAA but not produced in significant enough amounts by chlorination of DOC to prevent its use as the surrogate. 2-bromopropionic acid is used as the surrogate compound because the surrogate (2,3-dibromopropionic acid) listed in USEPA Method 552.2 coelutes with dibromochloroacetic acid on the GC column (Rtx-5, 5 percent diphenyl/95-percent dimethyl polysiloxane). A stock solution of 2-bromopropionic acid in MTBE at a certified concentration of 1 milligram per milliliter (mg/mL) is used to make the working surrogate solution in MTBE at a concentration of 10 micrograms per milliliter (μg/mL) by diluting 100-microliter stock solution with MTBE to a final volume of 10 mL using a volumetric flask and gas-tight syringe. Twenty microliters (μL) of the working surrogate solution are added to the sample just prior to extraction of the HAA to give 2-bromopropionic acid a concentration of 5 μg/L.
An internal standard (IS) that elutes with the methylated HAA during GC analysis is added to the samples and the HAA calibration standards. The measured peak areas for the analytes are normalized to the peak area of the IS to compensate for any small differences in GC injection volume or matrix effects on sample volitization in the injector between samples. 2-bromo-1-chloropropane (CH2ClCHBrCH3) is used as the IS because the IS (1,2,3-trichloropropane) listed in USEPA Method 552.2 coelutes with bromochloroacetic acid on the Rtx-5 column. An IS stock solution of 2-bromo-1-chloropropane in methanol is purchased at a certified concentration of 2 mg/mL. From this stock standard solution, the working IS solution in MTBE, at a concentration of 10 μg/mL, is prepared by diluting 50-microliter stock solution with MTBE to a final volume of 10 mL using a volumetric flask and gas-tight syringe. Ten μL of the working IS solution are added to every vial just prior to GC analysis to give an IS concentration of 0.4 μg/mL in the extract.
A primary stock standard solution containing all nine HAA in MTBE is purchased and used to prepare the working standards used for calibration. The stock standard solution is stored at –10°C and protected from light. It is stable for at least 1 month but should be checked for signs of evaporation. When purchasing commercially prepared standards, solutions prepared in methanol must not be used because the HAA are subject to spontaneous methylation when stored in this solvent (Xie and others, 1993). Furthermore, tribromoacetic acid is unstable in methanol because it undergoes decarboxylation when stored in this solvent (USEPA Method 552.2). The Supelco mix for USEPA Method 552.2 contains a mix of nine acids in the concentrations listed in table 2.
Table 2 (View this table on a separate page.)Haloacetic acid and surrogate concentrations in the stock standard solution and the three working standard solutions C, B, and A that are prepared in methyl-tert-butyl ether.
[Certificate of analysis accompanying stock standard solution lists actual concentration of haloacetic acid compounds to four significant figures. Because the actual concentrations vary slightly between lots, nominal concentrations in micrograms per milliliter are listed here instead. —, stock standard solution does not contain the surrogate compound]
AnalyteStock standard solution (μg/mL)Working standard solution C (μg/mL)Working standard solution B (μg/mL)Working standard solution A (μg/mL)
Bromochloroacetic acid400200202
Bromodichloroacetic acid400200202
Dibromochloroacetic acid1000500505
Dibromoacetic acid200100101
Dichloroacetic acid600300303
Monobromoacetic acid400200202
Monochloroacetic acid600300303
Tribromoacetic acid2000100010010
Trichloroacetic acid200100101
2-bromopropionic acid (surrogate)500505
The working standard solutions, C, B, and A, are prepared by combining the stock standard solution and surrogate stock solution to give the concentrations listed in table 2. Working standard solution C is prepared first by combining 250 μL of the stock standard solution and 250 μL of the surrogate stock solution to give a final volume of 500 μL in a vial. The working standard solution B is prepared by diluting working standard solution C by a factor of ten with MTBE. One hundred μL of working standard solution C are measured with a gas-tight syringe and diluted to a final volume of 1 mL in a volumetric flask. The working standard solution A is prepared by diluting working standard solution B by a factor of ten with MTBE. One hundred μL of working standard solution B are measured with a gas-tight syringe and diluted to a final volume of 1 mL in a volumetric flask.
These working standard solutions are used to prepare procedural calibration standards, which comprise nine concentration levels of each analyte, with the lowest standard being at or near the method detection limit (MDL) of each analyte.
The quality-control standard solution (QCSS) is an independent standard solution used to prepare quality-control samples to verify the calibration standards (see “Quality-Control Samples” section of this report). The QCSS is prepared by weighing 0.05 grams (g) of each neat compound into individual 5-mL volumetric flasks and diluting to volume with MTBE. The resulting concentration of these stock standards is 10,000 mg/L. Forty μL of each stock standard solution is then mixed and the combined solution diluted to a final volume of 2 mL with MTBE. This solution, QCSS A, contains 200 mg/L of each of the nine HAA compounds. QCSS B is prepared by diluting QCSS A by a factor of ten (200-μLQCSS A to a final volume of 2 mL with MTBE) to give a final concentration of 20 mg/L for each of the nine HAA compounds.
Water samples for HAAFP analysis should be collected using baked glass, Teflon, or stainless steel sampling containers. For example, shallow surface-water grab samples can be collected directly into baked amber glass bottles, and deeper surface-water-integrated samples can be collected with Teflon or stainless steel Van Dorn-type samplers for transfer into baked amber glass bottles. Exposure to organic solvents must be avoided. If the sampler is cleaned with methanol, copious amounts of water must be used to rinse the sampler to ensure that the methanol is removed completely prior to collecting the sample.
Water samples must be filtered prior to analysis of the HAAFP. Samples are filtered in the field or laboratory within 24 hours of collection. Procedures for collecting and filtering samples, such as those given in Chapters A4 and A5 of the National Field Manual for the Collection of Water-Quality Data (Radtke and others, 2002), can be used if modified to avoid contact between the sample and solvents or plastics. No preservatives are added to the samples. Each sample is assigned a unique number as it is logged into the Laboratory Information Management System (LIMS) (LabWorks, Analytical Automation Specialists, Inc.) and is stored at 4°C (39°F) until analyzed.
The procedure to form HAA by chlorination of water samples is the same as the procedure to form THM in the method for determination of THMFP (Crepeau and others, 2004). The procedure is summarized briefly here. The DOC and NH3-N concentrations of the sample are measured and used to calculate the appropriate amount of Cl2 used to dose the samples. Samples with a DOC concentration of 3 mg/L or greater usually are diluted with organic-carbon-free water prior to chlorination. If both HAAFP and THMFP are being determined for the sample, the same dilution factor is used for both analyses. A dose solution containing 3,000 to 4,000 mg/L residual-free Cl2 derived from sodium hypochlorite (NaOCl) and buffered to a pH of 8.3 with 1 molar (M) H3BO3 and 0.11 M NaOH is prepared. The pH of the sample is adjusted to a range between 8.3 and 8.7 by addition of dilute NaOH or hydrochloric acid (HCl). The sample is poured into three 40-mL amber glass vials with Teflon-faced septa, and sufficient dosing solution is added to satisfy the Cl2 demand of the DOC and NH3-N and to leave a residual-free Cl2 concentration of 2–4 mg/L after the incubation period. The vials are sealed headspace-free and incubated at 25°C (77°F) in the dark for 7 days. After incubation, one vial is opened to measure the pH and free Cl2. The pH must be 8.3 ± 0.1 and the residual-free Cl2 must be between 2 and 4 mg/L. If these parameters are not met, then the sample is redosed and incubated for another 7 days. The remaining two vials are quenched by adding sufficient Na2SO3 solution to neutralize the residual-free Cl2.The samples are refrigerated and can be held up to 14 days before extraction.
The samples are removed from refrigeration and allowed to equilibrate to room temperature. Two aliquots of every sample are analyzed, one undiluted and one diluted 1:5 with organic-free water. For the undiluted sample aliquot, one quenched vial is opened and 40 mL of sample water is measured with a graduated cylinder (which has been calibrated “to deliver” at 20°C with a 1-percent tolerance) and poured into a precleaned 60-mL vial with a Teflon-lined screw cap. For the diluted sample aliquot, the second quenched vial is opened, and 8 mL of sample are pipetted into a 60-mL vial along with 32 mL of organic-free water. The final concentration result for each compound in a sample is derived only from analysis of the diluted aliquot if the concentration measured in the undiluted aliquot is higher than the concentration in the highest standard. Twenty μL of the 10.0 μg/mL 2-bromopropionic acid surrogate solution is added to every 60-mL vial. When adding surrogate or standard solutions to aqueous samples, the tip of the syringe must be well below the water level to avoid loss by volatilization. After injection of the surrogate solution, the sample vial is capped immediately and inverted to ensure mixing of solutions. The pH is adjusted to less than 0.5 by adding 2 mL of concentrated H2SO4. Two grams (g) of copper II sulfate pentahydrate and 16 g of Na2SO4 are added immediately to the sample using the heat produced from the H2SO4 addition to help dissolve the salts. The samples are shaken until the salts are dissolved (approximately 2–3 minutes). Then, 2.0 mL of MTBE are added to the samples and they are shaken vigorously for 2 minutes. The MTBE and aqueous layers are allowed to separate for approximately 5 minutes.
A disposable glass pipette is used to remove as much of the MTBE layer (upper) as possible (minimum of 1 mL) and to place it into a 4-mL silanized autosampler vial that will be used as the reaction vessel for the methylation step. Then, 0.5 mL of 10 percent (volume/volume) H2SO4 in methanol is added to each autosampler vial. The cap is tightened securely and the vials are placed in a water bath at 50°C (122°F) for 2 hours to allow for methylation of the analytes.
The vials are removed from the water bath and placed in an ice bath for 5 minutes. Two mL of 0.1 normal (N) NaOH is added to each vial and the vials are shaken for approximately 2 minutes. The MTBE and aqueous layers then are allowed to separate.
A second set of autosampler vials containing silanized glass inserts is used to hold the final samples for GC analyses. Ten μL of the IS solution (10 μg/mL 2-bromo-l-chloropropane) is placed in the silanized insert just prior to addition of the sample extract. Exactly 250-μL of the MTBE layer (upper) is transferred into the insert using a 250-μL fixed volume micropipettor. The vials are capped and shaken or mixed with a vortex mixer.
The samples should be analyzed on the GC as soon as possible after preparation because the final extract solutions deteriorate after a few days. Losses of dibromoacetic acid, dibromochloroacetic acid, and tribromoacetic acid in particular were observed. If the sample must be analyzed on the GC for a second time, it is recommended that a new aliquot of the water sample be processed rather than reanalyzing the old final extract samples.
This method for extraction and methylation uses only one-half the volume of MTBE that is required for USEPA Method 552.2 (U.S. Environmental Protection Agency, 1995). Therefore, the volume of excess final extract that must be disposed of as solvent waste is much less.
The instrument consists of a 6890 Hewlett Packard autosampler connected to a 5890 Hewlett Packard GC equipped with an Rtx-5 capillary column and an ECD. The autosampler is set to deliver 1-μL samples to the injection port of the GC. The GC operating configuration is summarized in table 3. Figure 1A illustrates the performance on the Rtx-5 column with the method analytes at standard level 5 (see table 5 for concentrations), an extraction blank, and a full procedural blank. The peaks for all nine HAA, the IS and the surrogate compound are well resolved. Baseline separation is achieved for all peaks, except for trichloroacetic acid [retention time (RT) = 10.745 min] and bromochloroacetic acid (RT = 10.864 min), which nearly are baseline resolved. The identities of the peaks with retention times of 6.333 min and 9.479 min are not known. However, because these extra peaks are not present in chromatograms for full procedural blanks (fig. 1C), they likely are due to contaminants in the standards rather than contaminants introduced during sample preparation. The Rtx-5 column was chosen instead of the DB-5.625 column used in USEPA Method 552.2 because the response for the monochloroacetic acid was better, as was the general performance of the column.
Table 3 (View this table on a separate page.)Gas chromatograph with electron capture detector components and specifications.
ComponentsSpecifications
ColumnRtx-5 30-meter x 0.25 millimeter internal diameter with 0.25 micrometer film thickness
Carrier gasNitrogen at 1 milliliter per minute flow at 40ºC
Oven40ºC for 15 minutes,
 40-110ºC at 7ºC/minute,
 110-250ºC at 20ºC/minute
InjectorSplit 10:1, 200ºC
DetectorElectron capture at 300ºC
Calibration standards are prepared using the same extraction and methylation procedures as for water samples. Nine calibration standards are made by adding appropriate volumes of the working standard solutions A, B, and C to 40-mL aliquots of fortified organic-free water (table 4). The organic-free water is fortified with Cl2 dosing solution and sodium sulfite quenching solution in approximately the same concentrations used for generating the HAA in water samples. These solutes may affect extraction efficiency of the HAA and, thus, should be present in the same concentrations in standards and samples. Three working standard solutions (A, B, and C) are used so that the volume of MTBE added to the calibration standard solutions is less than 20 μL for all nine calibration standards. The concentration of the nine HAA compounds and the surrogate compound in the nine calibration standards are listed in table 5.
Table 4 (View this table on a separate page.)Volumes of working standard solutions added to 40-milliliter samples of fortified organic-free water to make nine standard levels.
[µL, microliters]
Standard levelWorking standard solutionVolume (μL)
1A4
2A8
3A20
4B4
5B8
6B20
7C4
8C8
9C20
Table 5 (View this table on a separate page.)Concentrations of haloacetic acids and surrogate compound in nine standard levels.
[Std, standard. µg/L, micrograms per liter]
AnalyteStd level 1 (μg/L)Std level 2 (μg/L)Std level 3 (μg/L)Std level 4 (μg/L)Std level 5 (μg/L)Std level 6 (μg/L)Std level 7 (μg/L)Std level 8 (μg/L)Std level 9 (μg/L)
Bromochloroacetic acid0.20.41.02.04.0102040100
Bromodichloroacetic acid0.20.41.02.04.0102040100
Dibromochloroacetic acid0.51.02.55.0102550100250
Dibromoacetic acid0.10.20.51.02.05.0102050
Dichloroacetic acid0.30.61.53.06.0153060150
Monobromoacetic acid0.20.41.02.04.0102040100
Monochloroacetic acid0.30.61.53.06.0153060150
Tribromoacetic acid1.02.05.0102050100200500
Trichloroacetic acid0.10.20.51.02.05.0102050
2-bromopropionic acid (surrogate)0.51.02.55.0102550100250
The calibration curves are generated using the IS technique. Peak areas for the nine HAA compounds and the surrogate compound are normalized to the peak area of the IS in the same injection. Because the IS is presented in the same concentration in all of the final extracts analyzed on the GC, this normalization compensates for any small variations in GC injection volume or differences in matrix effects in the final extract solutions between samples.
The calibration curves for all nine HAA and the surrogate compound are quadratic. The coefficient of determination, R2, is used to assess the fit between each quadratic equation and the data for each analyte from the nine standard solutions. The R2 value must be 0.9980 or better or a new calibration curve must be generated.
Samples are analyzed immediately after extraction and methylation. The order of analysis begins with an MTBE instrument blank to verify that the instrument is free of contamination. Next, the standard curve is produced by analyzing the nine standards from lowest to highest concentrations, followed by another MTBE instrument blank. Then, the extraction blanks and samples are analyzed, followed by two continuing calibration verification standards and another MTBE instrument blank between sets of samples (see “Quality-Control Practices” section of this report). The final data for a sample may be combined from two or more different analyses where the data for each HAA analyte are taken from the dilution that fits within the standard curve.
The HAAFP data are processed using EZChrom chromatography software and archived in the USGS California Water Science Center’s LIMS. Data for selected samples also are entered into the USGS National Water Information System (NWIS) database.
In each of the samples, the software identifies the peaks of the nine HAA species, the IS, and the surrogate by their retention times and then converts measured peak areas to concentrations by normalizing the peak areas to the peak area of the IS and then converting normalized peak areas to concentrations using the standard calibration curves. The retention time for each compound must be within a 0.25-minute window of the expected value from the calibration curve. Full separation of the compounds is achieved with the Rtx-5 column at the analyzed concentrations (fig. 1). The analyst examines the chromatograms to verify that the peak identifications are correct. Compound interferences on the column are minimal due to the selectiveness of the extraction method. The EZChrom software automatically flags samples if the surrogate concentrations, reproducibility of the duplicate samples, or concentrations of the calibration verification standards are out of acceptable range (see "Quality Control Practices” section of this report). The analyst also examines these data. The individual chromatograms, calibration curve information, and quality-control data are archived and the archived site is linked to the LIMS.
The data are first imported from EZChrom into a spreadsheet (Microsoft Excel) for verification and calculations. After all quality-control criteria for a set of samples are met satisfactorily, the data are transferred to the LIMS. The data are accessible to users of the LIMS after the analyst verifies the final concentrations.