The Chemical Analysis of Argonne Premium Coal Samples

Edited by Curtis A. Palmer

1997
U.S. Geological Survey Bulletin 2144


Determination of Mercury and Selenium in Eight Argonne Premium Coal Samples by Cold-Vapor and Hydride-Generation Atomic Absorption Spectrometry

By Richard M. O'Leary

Table 1. Operating conditions for determination of mercury and selenium by cold-vapor absorption spectrometry (CVAAS) and hydride-generation AAS (HGAAS), respectively.

  Hg Se
Source Hg hollow
cathode lamp.
Electrodeless discharge lamp (EDL).
Slit 0.7 nm 0.7 nm
Wavelength 253.7 nm 196.0 nm
Quartz cell temperature 100°C 2,000°C
Mode Absorbance Absorbance
Recorder 2 mV, 5 mm/
min
10 mV, 5 mm/
min
AA recorder mode TC3 TC3
Sensitivity (peak height) 1 µg/L = 19 mm 10 µg/L = 72 mm