U.S. Geological Survey Open-File Report 01-257
By Kinga M. Revesz,
Jurate M. Landwehr, and Jerry Keybl
APPENDIX A: COMPLETE STEP BY STEP PROCEDURE
Weighing and Storing Samples:
- Samples should be dried overnight in a 90ºC oven and kept in sealed vessels.
- Condition balance (this step should be done once a day):
a. |
Place empty aluminum boat on balance
and close door. Wait for reading to stabilize (the 'g' on the display
appears). Tare the balance. |
b. |
Remove and replace boat and make
sure the stabilized weight is 0.000 mg. |
c. |
Repeat these steps until the balance
is stable. |
- Remove boat from balance, add sample and weigh the filled boat. Repeat
until you have 400±20 µg.
- Quantitatively transfer sample into a dried sample vessel and seal.
- Mark sample ID and weight on the side of the vessel with a permanent marker.
- Weigh boat again. It should read 0.000 mg. If it doesn't, there is a problem
with the way in which the sample was weighed or transferred.
- Repeat steps 2-5 for every sample.
Note:
- Do not cross contaminate samples:
a. |
Make sure aluminum boat always weighs
0.000 mg when empty. |
b. |
Make sure the spatula and sample
area are cleaned using chem wipes. |
- Always allow balance to stabilize before removing boat.
- Use the same boat until it becomes unusable.
Washing Sample Vessels:
- Remove marker with chem wipe and Propanol (isopropyl alcohol).
- Remove and dispose of the caps; place vessels in tray.
- Rinse set of vessels 8 times with tap water and once with de-ionized (DI)
water. Use mesh to hold vessels in place when emptying them.
- Fill ultrasound with DI water. Place the vessels in the DI water.
- Ultrasound the vessels for 30 minutes.
- Remove the vessels and empty the water out of them. Place the vessels in
a tray.
- Dry vessels in a 90ºC oven for at least 24 hours.
- Check the seals on the new caps and lay them out to cover the dried vessels
with.
- Remove vessels one by one from oven; cap dried vessels immediately to keep
out moisture.
Loading Samples in Tray:
- The needle assembly (sampling arm) must be moved out of the way by the GasBench
II controller located on the left side of arm.
- At job queue screen, press F1 (MENU).
- Turn knob until SETUP is highlighted. Press 'Enter' (button is located inside
the knob).
- Turn knob until OBJECTS is highlighted. Press 'Enter.'
- Turn knob until TRAYS is highlighted. Press 'Enter.'
- Turn knob until 2ml-I is highlighted. Press 'Enter.'
- Turn knob until OFFSET X (-44.2mm) is highlighted. Press 'Enter.'
- Remove sample bath cover.
- Load samples using the left upper port as 1. Note port, sample ID, and
weight as the samples are entered. The sequence is loaded left to right, top
to bottom.
- Place vessels containing 100% pure phosphoric acid (H3PO4) into the last
3 ports.
- Replace bath cover. (Make sure the hole for the cable is located on the
backside of the cover.)
- Press F4 (HOME) to return the sampling arm to its normal position.
Flush Samples:
- Make sure that the flushing needle is in the GasBench II sampling arm.
- Open the valve of the flushing helium tank. Check the flow of the GasBench
II regulator marked "Ref 3." It should be set at 21 psi.
- ISODAT Settings:
a. |
Select GASBENCH from CONFIG column.
|
b. |
Select "A-CON-B" from ENVIR. Column |
c. |
Select "2. SEQUENCE ACQ." (If
the option is not visible, hit ALT-T.) |
d. |
Make sure TABLE NAME value is highlighted. |
e. |
Select TABLE DIR from status bar
at bottom of screen. |
f. |
Select FLUSH from options and right-click
it. |
g. |
Check to make sure START LINE
and END LINE match what samples you have loaded. |
h. |
Click SAVE from status bar. |
i. |
Click MEASURE from status bar to
begin flushing. |
Add Sample Information to Queue:
- Select GASBENCH from CONFIG column.
- Select EDIT-B from ENVIR. Column.
- Select 2. SEQUENCE EDITOR. If you do not see it, hit ALT-T.
- Select CACO3.
- Click EDIT from status bar.
- Enter GasBench II Port # / GasBench II Method #
- Enter Sample ID in SAMPLE IDENT field. Enter must be hit after every change
in order for ISODAT to accept it.
- Enter process File.
- Enter the mass into the SAMPLE SIZE field.
- Add any comments concerning the sample that you may have.
- Click NEXT LINE from status bar for the next sample.
- Repeat steps 6-9 for every sample.
- When finished, click EXIT/SAVE from status bar.
Note:
The EDIT MASK and COPY MASK commands can be used to avoid typing repetitive
data.
Add Acid:
- Move sample arm out of the way (see Loading Samples in Tray, steps
2-7).
- Place the gas tight syringe into the acid and suck up 0.1 ml of acid.
- Remove the syringe from the acid and wipe the extra acid of the tip with
a chem wipe.
- Punch the syringe into the rubber septum of the sample vessels. Hold the
plunger in place. It is best to punch the septum off center.
- Expel the acid and wait 30 seconds to make sure that no acid remains on
the needle. Remove the syringe.
- Repeat steps 2-5 for each sample vessel in the set.
- Press F4 (HOME) to return the sampling arm to its normal position.
- Wait 24 hours before running the samples.
Run Samples:
- Select GASBENCH from CONFIG column.
- Select ACON-B from ENVIR. Column.
- Select "2. SEQUENCE ACQ." If this option is not visible, hit ALT-T.
- Make sure the TABLE NAME field is highlighted. Select TABLE DIR from status
bar.
- Select CACO3 and right click on it.
- Change START LINE and END LINE to correspond to what samples will be run.
- Select SAVE from the status bar.
- Select MEASURE from the status bar.
Retrieve Data from ISODAT:
- Select GASBENCH and EVAL-B from the CONIG AND ENVIR. Columns.
- Select "1. OFFLINE EVALUATION." If this option is not visible, hit ALT-T.
- Select Sort Index (default is Spec #).
- Choose DIRECTORY from the status bar.
- Tag all the runs using the TAG ALL command and parenthesis.
- Select ONL-EDIT from the Screen. Column.
- Page down to the fourth page.
- Enter "NO" in the LIST TYPE field.
- Enter "ALL" in the PEAK REPORT field.
- Enter "a:\mmddyy.wks" in the FILE NAME field.
- Enter "FORMAT1" in the LIST TYPE field.
- Enter "LOT" in the FILETYPE field.
- Choose EVAL.
- When the exporting process is complete, choose EXIT.
Store Data in LIMS:
- Open file in Excel. (*.wks file from floppy)
- Select filter from the data menu. Filter runs 1, 2, 3, 11, 12, and 13.
- Fill the R column with asterisks, once the runs are filtered.
- Save the modified file.
- Start LIMS.
- Choose "Import Analysis."
- Choose Mass Spec Delta-XL.
- Click Import.
- Select the file that will be imported.
- Select columns F30 and F31. Check the boxes to import these two columns.
- Click Import.
Continue to
Appendix B: ISODAT "CAC03" Method
, or return to
Table of Contents