Evaluation of the Ott Hydromet Qliner for Measuring Discharge in Laboratory and Field Conditions

By Jason C. McVay

Thumbnail of and link to report PDF (2.55 MB)Abstract

The U.S. Geological Survey, in collaboration with the University of Iowa IIHR – Hydroscience and Engineering, evaluated the use of the Ott Hydromet Qliner using laboratory flume tests along with field validation tests. Analysis of the flume testing indicates the velocities measured by the Qliner at a 40-second exposure time results in higher dispersion of velocities from the mean velocity of data collected with a 5-minute exposure time. The percent data spread from the mean of a 100-minute mean of Qliner velocities for a 40-second exposure time averaged 16.6 percent for the entire vertical, and a 5-minute mean produced a 6.2 percent data spread from the 100-minute mean. This 16.6 percent variation in measured velocity would result in a 3.32 percent variation in computed discharge assuming 25 verticals while averaging 4 bins in each vertical. The flume testing also provided results that indicate the blanking distance of 0.20 meters is acceptable when using beams 1 and 2, however beam 3 is negatively biased near the transducer and the 0.20-meter blanking distance is not sufficient. Field testing included comparing the measured discharge by the Qliner to the discharge measured by a Price AA mechanical current meter and a Teledyne RDI Rio Grande 1200 kilohertz acoustic Doppler current profiler. The field tests indicated a difference between the discharges measured with the Qliner and the field reference discharge between -14.0 and 8.0 percent; however the average percent difference for all 22 field comparisons was 0.22, which was not statistically significant.

First posted April 1, 2015

For additional information, contact:
Director, Iowa Water Science Center
U.S. Geological Survey
P.O. Box 1230
Iowa City, IA 52244
http://ia.water.usgs.gov/

Part or all of this report is presented in Portable Document Format (PDF). For best results viewing and printing PDF documents, it is recommended that you download the documents to your computer and open them with Adobe Reader. PDF documents opened from your browser may not display or print as intended. Download the latest version of Adobe Reader, free of charge. More information about viewing, downloading, and printing report files can be found here.


Suggested citation:

McVay, J.C., 2015, Evaluation of the Ott Hydromet Qliner for measuring discharge in laboratory and field conditions: U.S. Geological Survey Open-File Report 2015–1052, 13 p., https://dx.doi.org/10.3133/ofr20151052.

ISSN 2331-1258 (online)



Contents

Acknowledgments

Abstract

Introduction

Purpose and Scope

Methods

Laboratory Results

Field Results

Summary and Conclusions

References Cited


Accessibility FOIA Privacy Policies and Notices

Take Pride in America logo USA.gov logo U.S. Department of the Interior | U.S. Geological Survey
URL: http://pubsdata.usgs.gov/pubs/of/2015/1052/index.html
Page Contact Information: GS Pubs Web Contact
Page Last Modified: Wednesday, 07-Dec-2016 21:41:02 EST