  <dc:contributor>J. M. Axelrod</dc:contributor>
  <dc:creator>I. Adler</dc:creator>
  <dc:date>1955</dc:date>
  <dc:description>A fluorescent x-ray spectrographic method for the determination of thoria in rock samples uses thallium as an internal standard. Measurements are made with a two-channel spectrometer equipped with quartz (d = 1.817 A.) analyzing crystals. Particle-size effects are minimized by grinding the sample components with a mixture of silicon carbide and aluminum and then briquetting. Analyses of 17 samples showed that for the 16 samples containing over 0.7% thoria the average error, based on chemical results, is 4.7% and the maximum error, 9.5%. Because of limitations of instrumentation, 0.2% thoria is considered the lower limit of detection. An analysis can be made in about an hour.</dc:description>
  <dc:format>application/pdf</dc:format>
  <dc:language>en</dc:language>
  <dc:title>Determination of thorium by fluorescent x-ray spectrometry</dc:title>
  <dc:type>article</dc:type>