<?xml version='1.0' encoding='utf-8'?>
<oai_dc:dc xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/ http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
  <dc:contributor>Anon</dc:contributor>
  <dc:creator>Falah Thamir</dc:creator>
  <dc:date>1991</dc:date>
  <dc:description>Several problems with the Mariotte reservoir system were discovered when it was used to apply a constant water pressure as a boundary condition for a prolonged period. The constant-pressure boundary condition is required for some laboratory experiments to study water flow through porous media. The observed problems were caused by temperature and barometric-pressure fluctuations while the flow rates were very small and caused erroneous water flow-rate measurements. An alternative method was developed and used where the water pressure is controlled by regulating its level by using water-level sensing electrodes. The new method eliminated the effects of temperature and barometric-pressure fluctuations and maintained an acceptable accuracy of the estimated water flow rate without compromising the advantages of the Mariotte reservoir.</dc:description>
  <dc:format>application/pdf</dc:format>
  <dc:language>en</dc:language>
  <dc:publisher>Publ by ASCE</dc:publisher>
  <dc:title>Alternative method to Mariotte reservoir system for maintaining constant hydraulic pressure</dc:title>
  <dc:type>text</dc:type>
</oai_dc:dc>