  <dc:contributor>J. Gordon</dc:contributor>
  <dc:creator>N. Latysh</dc:creator>
  <dc:date>2004</dc:date>
  <dc:description>A study was undertaken to investigate differences between laboratory and field pH measurements for precipitation samples collected from 135 weekly precipitation-monitoring sites in the National Trends Network from 12/30/1986 to 12/28/1999. Differences in pH between field and laboratory measurements occurred for 96% of samples collected during this time period. Differences between the two measurements were evaluated for precipitation samples collected before and after January 1994, when modifications to sample-handling protocol and elimination of the contaminating bucket o-ring used in sample shipment occurred. Median hydrogen-ion and pH differences between field and laboratory measurements declined from 3.9 ??eq L-1 or 0.10 pH units before the 1994 protocol change to 1.4 ??eq L-1 or 0.04 pH units after the 1994 protocol change. Hydrogen-ion differences between field and laboratory measurements had a high correlation with the sample pH determined in the field. The largest pH differences between the two measurements occurred for high-pH samples (&gt;5.6), typical of precipitation collected in Western United States; however low- pH samples (&lt;5.0) displayed the highest variability in hydrogen-ion differences between field and laboratory analyses. Properly screened field pH measurements are a useful alternative to laboratory pH values for trend analysis, particularly before 1994 when laboratory pH values were influenced by sample-collection equipment.</dc:description>
  <dc:format>application/pdf</dc:format>
  <dc:identifier>10.1023/B:WATE.0000022971.59349.fc</dc:identifier>
  <dc:language>en</dc:language>
  <dc:title>Investigation of differences between field and laboratory pH measurements of national atmospheric deposition program/national trends network precipitation samples</dc:title>
  <dc:type>article</dc:type>