Method of Analysis at the U.S. Geological Survey California Water Science Center, Sacramento Laboratory – Determination of Haloacetic Acid Formation Potential, Method Validation, and Quality-Control Practices
Barbara C. Zazzi, Kathryn L. Crepeau, Miranda S. Fram and Brian A. Bergamaschi
Table 8
Sequence of blanks, calibration standards, quality-control standards, and unknown samples analyzed during one run of the gas chromatograph.
[
MTBE, methyl
tert-butyl ether; Std, standard;
QCS, quality-control sample;
µg/L, micrograms per liter;
HAA, haloacetic acid;
CCV, continuing calibration verification]
| Vial number | Sample ID | Sample description |
|---|
| 1 | MTBE | Instrument blank |
| 2 | Std 1 | Calibration standards |
| 3 | Std 2 | Calibration standards |
| 4 | Std 3 | Calibration standards |
| 5 | Std 4 | Calibration standards |
| 6 | Std 5 | Calibration standards |
| 7 | Std 6 | Calibration standards |
| 8 | Std 7 | Calibration standards |
| 9 | Std 8 | Calibration standards |
| 10 | Std 9 | Calibration standards |
| | | |
| 11 | MTBE | Instrument blank |
| 12 | Extraction blank | Organic-free water carried through all extraction steps |
| 13 | Full procedural blank | Organic-free water carried through full procedure |
| 14 | QCS5 | Quality-control standard containing 5 µg/L of each HAA |
| 15 | QCS50 | Quality-control standard containing 50 µg/L of each HAA |
| 16 | MTBE | Instrument blank |
| 17 | Sample 1 diluted | Sample used for matrix spike, sample is diluted 1:5 with organic-free water |
| 18 | Matrix spike | 1:4 diluted sample 1 spiked with std 5 concentrations of HAA |
| 19 | Matrix spike duplicate | 1:4 diluted sample 1 spiked with std 5 concentrations of HAA |
| 20 | Sample 2 | Sample |
| | | |
| 21 | Sample 3 | Sample |
| 22 | Sample 4 | Sample |
| 23 | Sample 5 | Sample |
| 24 | Sample 6 | Sample |
| 25 | Sample 7 | Sample |
| 26 | Sample 8 | Sample |
| 27 | Sample 9 | Sample |
| 28 | Sample 10 | Sample |
| 29 | MTBE | Instrument blank |
| 30 | CCV – std 4 | Continuing calibration verification-std level 4 |
| | | |
| 31 | CCV – std 5 | Continuing calibration verification-std level 5 |
| 32 | MTBE | Instrument |
| 33 | Sample 1 | Same samples as in vials 17−29, but diluted 1:5 with organic-free water prior to extraction |
| 34 | Sample 2 diluted | Diluted sample |
| 35 | Sample 3 diluted | Diluted sample |
| 36 | Sample 4 diluted | Diluted sample |
| 37 | Sample 5 diluted | Diluted sample |
| 38 | Sample 6 diluted | Diluted sample |
| 39 | Sample 7 diluted | Diluted sample |
| 40 | Sample 8 diluted | Diluted sample |
| | | |
| 41 | Sample 9 diluted | Diluted sample |
| 42 | Sample 10 diluted | Diluted sample |
| 43 | MTBE | Instrument blank |
| 44 | CCV – std 4 | Continuing calibration verification–std level 4 |
| 45 | CCV – std 5 | Continuing calibration verification–std level 5 |
| 46 | MTBE | Instrument blank |