Table 3. Results of laboratory quality-assurance tests of autosamplers.

[Results in bold indicate potential contamination or carryover. Results in italics represent statistics or values not derived from the laboratory. Abbreviations: DI, distilled, deionized water; SRS, Standard Reference Sample; <, less than; N/A, not applicable]

Test sequence Sampling
method
Purpose Medium Replicate No. Total suspended solids (mg/L) Ammonium-nitrogen (mg/L as N) Total Kjeldahl nitrogen(mg/L as N) Total phosphorus (mg/L as P) Soluble reactive phosphorus (mg/L as P) Chloride (mg/L) Comments
1 Direct pour (no autosampler) Test of lab cleanliness DI 1 N/A < 0.01 < 0.05 < 0.025 0.009 < 0.25 Tests 1-3 are analysis of source materials, for comparison with results of samples collected with the autosampler in tests 4,6,7,8.
2 < 0.01 < 0.05 < 0.025 < 0.005 < 0.25
Median < 0.01 < 0.05 < 0.025 0.007 < 0.25
2 Direct pour (no autosampler) Test of lab accuracy SRS, low N/A N/A 0.608 0.292 0.073 0.038 5.6
Expected value N/A 0.592 0.592 0.072 0.037 5.3
3 Direct pour (no autosampler) Test of lab accuracy SRS, high N/A N/A 1.03 0.732 0.127 0.061 6.2
Expected value N/A 0.986 0.986 0.156 0.051 5.9
4 Autosampler Test of initial cleaning DI blank 1 N/A < 0.01 < 0.05 < 0.025 < 0.005 < 0.25  
2 < 0.01 < 0.05 < 0.025 0.007 < 0.25
3 < 0.01 < 0.05 < 0.025 0.007 < 0.25
Median < 0.01 < 0.05 < 0.025 0.007 < 0.25
Standard Deviation 0.000 0.000 0.000 0.004 0.0
5 Autosampler Establish dirty tubing Water + dirt 1 61 0.299 0.837 0.306 0.139 3.6 Quantity of dirt undefined. Sampled from bucket, mixed by hand.
2 57 0.297 0.976 0.301 0.132 3.6
3 48 0.299 0.96 0.294 0.133 3.6
Median 57 0.299 0.96 0.301 0.133 3.6
Standard Deviation 6.7 0.001 0.076 0.006 0.004 0.0
6 Autosampler Test of cross contamination DI blank 1 < 0.2 < 0.01 0.055 < 0.025 < 0.005 < 0.25 Tests carryover from high TSS samples in Test 5.
2 < 0.2 < 0.01 < 0.05 < 0.025 < 0.005 < 0.25
3 N/A < 0.01 < 0.05 < 0.025 < 0.005 < 0.25
Median <0.2 <0.01 < 0.05 <0.025 <0.005 <0.25
Standard Deviation 0.0 0.000 0.032 0.000 0.000 0.0
7 Autosampler Test of dilution SRS, high 1 N/A 1.02 0.943 0.141 0.053 6.2 Tests high SRS and possibility of dilution when sampling low concentration sources before high concentration sources.
2 1.02 0.947 0.138 0.054 6.2
3 1.04 0.964 0.143 0.055 6.2
Expected value 0.986 0.986 0.156 0.051 5.9
Median 1.02 0.947 0.141 0.054 6.2
Standard Deviation 0.012 0.011 0.003 0.001 0.0
8 Autosampler Test of cross contamination SRS, low 1 N/A 0.616 0.591 0.069 0.039 5.6 Tests low SRS and possibility of cross contamination when sampling high-concentration sources before low concentration sources.
2 0.62 0.609 0.069 0.039 5.6
3 0.626 0.603 0.068 0.045 5.6
Expected value 0.592 0.592 0.072 0.037 5.3
Median 0.62 0.603 0.069 0.039 5.6
Standard Deviation 0.005 0.009 0.001 0.003 0.0

First posted June 18, 2010

For additional information contact:
Director, Oregon Water Science Center
U.S. Geological Survey
2130 SW 5th Ave.
Portland, Oregon 97201
http://or.water.usgs.gov

Part or all of this report is presented in Portable Document Format (PDF); the latest version of Adobe Reader or similar software is required to view it. Download the latest version of Adobe Reader, free of charge.

Accessibility FOIA Privacy Policies and Notices

Take Pride in America logo USA.gov logo U.S. Department of the Interior | U.S. Geological Survey
URL: http:// pubsdata.usgs.gov /pubs/sir/2010/5008/table3.html
Page Contact Information: Contact USGS
Page Last Modified: Thursday, 10-Jan-2013 19:11:41 EST