Techniques and Methods 5 A–8
Chapter 8 of
Book 5, Laboratory Analysis
Section A, Water Analysis
Part 1. Method for the Digestion, Oxidation, and Analysis of Total Mercury ...
Part 2. Method for the Oxidation and Analysis of Total Mercury in Suspended...
Table 1. Results from multiple analyses of solids for method detection limi...
Table 2. Total mercury concentrations for multiple analyses of clay substra...
Table 3. Percent recovery data for the clay substrate, organic substrate, a...
Table 4. Results from multiple analyses of particulate MESS-2 CRM for dete...
Table 5. Results for the analysis of MESS-2 CRM solution
Table 6. Results for the analysis of IAEA-405 CRM solution
Multiply | By | To obtain |
---|---|---|
Length | ||
centimeter (cm) | 3.94 x 10-1 | inch (in) |
millimeter (mm) | 3.94 x 10-2 | inch (in) |
micrometer (µm) | 3.94 x 10-5 | inch (in) |
Mass | ||
gram (g) | 3.53 x 10-2 | ounce, avoirdupois (oz) |
milligram (ng) | 3.53 x 10-5 | ounce, avoirdupois (oz) |
microgram (µg) | 3.53 x 10-8 | ounce, avoirdupois (oz) |
nanogram (mg) | 3.53 x 10-11 | ounce, avoirdupois (oz) |
Volume | ||
liter (L) | 2.64 x 10-1 | gallon (gal) |
liter (L) | 3.38 x 10-1 | ounce, fluid (oz) |
milliliter (mL) | 2.64 x 10-4 | gallon (gal) |
microliter (µL) | 2.64 x 10-7 | gallon (gal) |
Flow | ||
milliliters per minute (mL/min) | 6.10 x 10-2 | cubic inch per minute (in3/min) |
Pressure | ||
pounds per square inch (psi) | 6.89 | kilopascal (kPa) |
Resistivity | ||
megohm centimeter (MΩ-cm) | 3.94 x 10-1 | megohm inch (MΩ-in) |
Temperature in degrees Celsius (°C) may be converted to degrees Fahrenheit (°F) as follows:
°F=(1.8x°C)+32
Concentrations for suspended solids samples are in nanograms per liter (ng/L).
Concentrations for solids samples are in nanograms per grams (ng/g).
g | gram mg milligram (10-3 grams) |
µg | micrograms (10-6 grams) |
ng | nanograms (10-9 grams) |
L | liters |
mL | milliliters (10-3 liters) |
µL | microliters (10-6 liters) |
mg/L | milligrams per liter (parts per million) |
µg/L | micrograms per liter (parts per billion) |
ng/L | nanograms per liter (parts per trillion) |
ng/g | nanograms per gram (parts per billion) |
cm | centimeters (10-2 meters) |
mm | millimeters (10-3 meters) |
µm | micron (10-6 meters) |
mL/min | milliliters per minute |
MΩ-cm | megohm centimeter |
Mmol | mega moles (106 moles) |
Ar | argon |
BDL | batch detection limit |
BrCl | bromine monochloride |
CRM | certified reference material |
CVAFS | cold vapor atomic fluorescence spectroscopy |
DQO | data quality objective |
HCl | hydrochloric acid |
Hg | mercury |
Hg0 | gaseous mercury |
HgS | cinnabar |
HgT | total mercury |
HNO3 | nitric acid |
I.D. | inside diameter |
KBr | potassium bromide |
KBrO3 | potassium bromate |
MDL | method detection limit |
N2 | nitrogen |
NIST | National Institute of Standards and Technology |
NH2OH*HCl | hydroxylamine hydrochloride |
O.D. | outside diameter psi pounds per square inch |
QA | quality assurance |
QCS | quality-control sample |
QFF | quartz fiber filter |
SnCl2 | stannous chloride |
USEPA | U.S. Environmental Protection Agency |
USGS | U.S. Geological Survey |
WDML | Wisconsin District Mercury Lab |
The methods documented in this report are utilized by the Wisconsin District Mercury Lab for analysis of total mercury in solids (soils and sediments) and suspended solids (isolated on filters). Separate procedures are required for the different sample types. For solids, samples are prepared by room-temperature acid digestion and oxidation with aqua regia. The samples are brought up to volume with a 5 percent bromine monochloride solution to ensure complete oxidation and heated at 50°C in an oven overnight. Samples are then analyzed with an automated flow injection system incorporating a cold vapor atomic fluorescence spectrometer. A method detection limit of 0.3 ng of mercury per digestion bomb was established using multiple analyses of an environmental sample. Based on the range of masses processed, the minimum sample reporting limit varies from 0.6 ng/g to 6 ng/g. Suspended solids samples are oxidized with a 5 percent bromine monochloride solution and held at 50°C in an oven for 5 days. The samples are then analyzed with an automated flow injection system incorporating a cold vapor atomic fluorescence spectrometer. Using a certified reference material as a surrogate for an environmental sample, a method detection limit of 0.059 ng of mercury per filter was established. The minimum sample reporting limit varies from 0.059 ng/L to 1.18 ng/L, depending on the volume of water filtered.
Since the mid-1980's, scientists have redefined our understanding of the sources, cycling processes and pathways, and levels of mercury (Hg) in various environmental media. Much of this improved understanding has been facilitated through the advancement of sampling and analytical procedures that pointed to low-level atmospheric deposition as the key Hg source reaching all aquatic ecosystems across the globe. In addition, even though most soils, sediments, and suspended particulates in surface waters have been relatively enriched with Hg over time, concentrations are still quite low and remain a significant analytical challenge to quantify accurately. The dominant repository for the vast majority of Hg released to the environment over the past 100 years has been surficial soils and sediments (soils and sediments: 947 Mmol; atmosphere: 17 Mmol; oceans: 36 Mmol)(Mason and others, 1994). Thus, our ability to quantify and monitor these large pools of Hg is critical for assessing and continuing to improve our understanding of the global Hg problem. The precise chemical composition of Hg in soils, sediments, and suspended particulates remains difficult to quantify analytically, although in most instances it is inorganic, divalent Hg that is most likely sorbed to fine-grained, organic substrates (Wiener and others, 2003). Lacking species-specific methods, most researchers report total mercury (HgT) and methylmercury concentrations for solid-phase environmental samples (DeWild and others, 2004). At sites distant from point sources, most soil and sediment Hg is derived from atmospheric deposition, and HgT levels range from about ten to several hundred nanograms per gram (ng/g, dry weight) (Nater and Grigal, 1992; Krabbenhoft and others, 1999). Near point sources, especially cinnabar (HgS) deposits or abandoned placer mines, Hg concentrations can be considerably higher, generally in the part per million range (Krabbenhoft and others, 1999; Biester et al. 2000; and Gustin et al. 2000). With such a wide range of concentrations observable in soils and sediments, several factors must be considered when attempting to quantify HgT in soils, sediment, and suspended particulates, including: (1) sample contamination due to improper sample handling or Hg-contaminated sampling devices; (2) natural heterogeneity of native materials; and, (3) matrix interferences and incomplete oxidation of samples.
This document describes the methods used by the U.S. Geological Survey (USGS) Wisconsin District Mercury Laboratory (WDML) in Middleton, Wisconsin, to analyze solids and suspended solids samples for HgT concentration and to document a method detection limit (MDL) for these sample media using the described techniques. Because the WDML employs different procedures for the preparation of solids (soils and sediments) and for suspended solids that are collected (isolated) on filters, the procedures, MDL's, and method performance are presented separately in this report. In addition, although this document will not address the need to practice "trace-metal-clean" methods during sample collection (Olson and DeWild, 1999), the utmost care should be taken to ensure reliable results when acquiring environmental samples for trace Hg levels.
The WDML gratefully acknowledges support for this study from the USGS Toxic Substances Hydrology Program.
This method is suitable for the determination of HgT in solids collected from terrestrial and aquatic environments. Samples can consist of soils or bed sediments and can be processed wet or freeze-dried. The minimum mass of Hg that is detectable is 0.3 ng per digestion bomb. The upper range can be extended to meet most samples by controlling the amount of material processed and through dilutions of the digestate. All samples are reported in ng/g dry weight.
An aliquot of solid material homogenized with a Teflon policeman is digested and oxidized in a Teflon digestion bomb with aqua regia at room temperature overnight to convert all Hg to Hg2+. The digested sample is then diluted to volume with 5 percent bromine monochloride (BrCl). After dilution, the sample is pre-reduced with hydroxylamine hydrochloride (NH2OH*HCl) to remove any free halogens, then reduced with stannous chloride (SnCl2) to convert Hg2+ to gaseous mercury (Hg0). The Hg0 is purged, captured on a gold trap, thermally desorbed, and then detected using a cold vapor atomic fluorescence spectrometer (CVAFS). This method can be used to determine HgT concentrations in solid samples with a method detection limit of 0.3 ng in a digestion bomb.
HgT analysis is extremely sensitive to contamination; therefore, care must be taken to avoid contamination during sample collection and analysis. Sample collection should be conducted using clean hands/dirty hands protocol (Olson and DeWild, 1999). Solids samples are collected and placed into precleaned vials. Collection and analysis equipment is cleaned according to the procedures outlined in DeWild and others (2002). Vials can consist of Teflon, cleaned according to DeWild and others (2002), baked glass vials (prepared by heating to 550°C for 4 hours), or acid-rinsed polycarbonate vials. Samples are frozen as soon as possible after collection, shipped to the lab on dry ice by overnight mail, and held at a temperature of -15°C or less until processing. The WDML has not performed a holding time study; however, a frozen certified reference material (CRM) certified for Hg is available through the National Institute of Standards and Technology (NIST) and is stable for a duration of 9 years.
Although samples are reported in ng/g dry weight, the MDL represented as a mass is the most effective way for the WDML to evaluate detection. It is not possible to consistently weigh out 100 mg of solids for every sample, which is the target sample size the WDML uses. Because varying amounts of solids are processed, using an MDL based on mass per aliquot allows a direct comparison of the amount of mercury in the sample digestate to the consistent value of 0.3 ng. Although the target mass for sample processing is 100 mg, the actual mass varies from 50 mg to 500 mg. This operational mass range yields a minimum sample reporting limit of 0.6 ng/g to 6 ng/g, as processed.
To demonstrate that the WDML can accurately quantify HgT in solid samples at environmentally important levels, an MDL study was performed. Even though wet samples were used for this study, the MDL is applicable to both wet and freeze-dried solid samples. The MDL reflects the mass of Hg in a digestion bomb, and is used to evaluate the mass of Hg in samples regardless of their condition (wet or freeze-dried) at the time of digestion. An MDL of 0.3 ng of HgT was determined according to U.S. Environmental Protection Agency (USEPA) protocol (U.S. Environmental Protection Agency, 1990) from multiple analyses of a solid sample that was primarily clay (table 1). The sample was mixed thoroughly, and ten aliquots were transferred into Hg clean Teflon digestion bombs. The target mass of 100 mg was not achieved for each aliquot; therefore, the mass detected per bomb was converted to a 100 mg equivalency. The samples were analyzed over ten days.
Table 1. Results from multiple analyses of solids for method detection limit (MDL) assay.
[All masses in nanograms (ng) of mercury per 100 milligrams (mg) of solids]
Clay Substrate | |
---|---|
1.39 | |
1.53 | |
1.73 | |
1.55 | |
1.60 | |
1.40 | |
1.43 | |
1.20 | |
1.60 | |
1.42 | |
Average | 1.49 |
Standard deviation | 0.1 |
Percent relative standard deviation | 10.0 |
Detection limit (standard deviation x 2.821*) | 0.3 |
*students T-value at the 99 percent confidence interval for n=10
Upon receipt at the laboratory or on the day of preparation, reagent containers should be labeled with the date received or made and the initials of the person preparing them.
The stock and substock standards should by stored outside of the clean laboratory to prevent contamination of the entire lab. To clean a volumetric flask, fill to approximately 20 percent total volume with 50 percent HNO3, place the ground glass stopper on its side over the opening to prevent pressure buildup, and reflux on a hotplate for 4 hours.
The quality-control sample (QCS) will be prepared from a commercially available Hg standard. The standard will be of a different source or lot than the standard used to prepare the analytical standards. The QCS is used throughout analysis to verify statistical control.
Prior to analysis, solids samples are digested and oxidized to ensure complete conversion of Hg to Hg2+. Any number of samples may be prepared in a digestion batch. For every 10 environmental samples, there will be a method blank, a CRM sample, and one of the 10 environmental samples will be prepared in triplicate. A minimum of three method blanks must be created for a digestion batch. A method blank is an empty Teflon digestion bomb that has all the reagents added and is carried through all the steps that a sample would undergo.
Solids samples are completely thawed, then thoroughly homogenized using a Teflon policeman. A small aliquot, approximately 100 mg, is transferred to a Teflon digestion bomb using the policeman. The exact mass transferred is recorded to the nearest tenth of a mg, as well as the bomb ID. When all the solids are transferred to the digestion bombs, 6 mL of HCl and 2 mL of HNO3 are added. This combination of HCl and HNO3, in a 3:1 ratio, creates aqua regia, a strong oxidizing solution. The bombs are loosely capped and placed in a fume hood for one hour. After the hour has expired, the bombs are then wrench-tightened, and allowed to react overnight at room temperature. After 24 hours, the bombs are diluted with 22 mL of 5 percent BrCl to a total volume of 30 mL. When the samples have been diluted, they are wrench-tightened and placed in an oven at 50°C overnight, after which time they are ready for analysis.
After the digestion bombs are removed from the oven, the digestate is thoroughly mixed, an aliquot is transferred to a 60-mL Teflon vial containing 50 µL NH2OH*HCl, and diluted to 50 mL. These diluted samples are then analyzed according to USEPA Method 1631, Revision E (U.S. Environmental Protection Agency, 2002) utilizing an automated flow injection system. An analytical batch consists of 12 vials bracketed by the appropriate high level and low level QCS. Eleven vials contain samples, and the twelfth vial is a matrix spike of an environmental sample contained in one of the previous 11 vials. All samples are analyzed in duplicate from their respective vials and must fall within the calibration range. If they do not, they must be diluted further in a fresh 60-mL Teflon vial. A sample can consist of a method blank, a CRM, or an environmental sample.
Data Quality Objectives (DQOs) have been established to evaluate statistical control over the analytical system during the analysis.
where
RQCS = percent recovery for the QCSwhere
% Diff = Percent difference for duplicate analysiswhere
RSP = Percent recovery for matrix spikeThe following formulas are used to calculate concentrations.
Cng/L = (PA/S) * D, (4)
where
Cng/L = concentration of digestate, in ng/L
PA = peak area
S = slope of calibration line
D = dilution factor (50/sample aliquot volume in mL)
MBB = Cng/L * VB, (5)
where
MBB = mass of Hg in method blank, in ng
Cng/L = concentration of digestate, in ng/L, as determined in Equation 4
VB = volume of digestate originally in bomb, in L
where
MS = mass of Hg in sample bomb, in ng
Cng/L = concentration of digestate, in ng/L, as determined in Equation 4
VB = volume of digestate originally in bomb, in L
MBB–AVE = average mass in all method blanks for digestion batch
where
Cng/g = concentration of sample, in ng/g
MS = mass of Hg in sample bomb, in ng
WS = mass of solids in bomb, in g
Included with the digestion batch of environmental samples are method blanks, replicate analyses, and CRM samples. Each of these samples provides quality-control information used to evaluate the acceptability of the analytical runs. Typically, multiple analytical runs are necessary to analyze the entire digestion batch.
where
BDL = Batch detection limit σ = Standard deviation of the method blanks for the digestion batch, in ngTo evaluate method performance, three separate solids samples were analyzed over 10 days, representing different types of materials. One was a predominately clay substrate, the second was an organic rich substrate, and the third was a sandy, mineral substrate. Each sample was set up in 20 separate bombs, with half of those having a 1 ng Hg spike added to them to evaluate the digestion and oxidation processes.
Precision was evaluated by examining the percent relative standard deviation of the concentrations obtained from all analyses of the solid samples for each type of material. The percent relative standard deviation for the clay substrate was 10.0 and the organic rich substrate was 12.8 (table 2). The sand substrate was not used to evaluate precision because the concentration of the samples was less than the MDL. Percent relative standard deviation was calculated using the formula:
[All concentrations in nanograms per gram (ng/g) wet weight]
Clay Substrate | Organic Substrate | |
---|---|---|
13.9 | 12.6 | |
15.3 | 10.1 | |
17.3 | 12.5 | |
15.5 | 13.0 | |
16.0 | 11.4 | |
14.0 | 11.4 | |
14.3 | 11.9 | |
12.0 | 11.9 | |
16.0 | 16.0 | |
14.2 | 11.3 | |
Average | 14.9 | 12.2 |
Standard deviation | 1.5 | 1.6 |
Percent relative standard deviation | 10.0 | 12.8 |
where
% RSD = percent relative standard deviation
CSTDEV = standard deviation of all samples
CAVE = average concentration of all samples
Accuracy was evaluated by calculating the percent recovery for the matrix spikes versus the unspiked samples. All three substrates were used to evaluate accuracy because a known amount of Hg was added to the digestion bomb and a recovery could be based on that addition. Percent recoveries were calculated using the formula:
where
R = percent recovery
MS = analytically determined mercury mass in aliquot
CA = average concentration of unspiked sample
WS = mass of sample aliquot
SM = mass of spike added
Recoveries ranged from 75 to 176 percent with an average of 113 and a standard deviation of 31 for the clay substrate sample (table 3). For the organic substrate sample, only four samples could be evaluated for accuracy due to operational problems encountered. The identification codes were in question for six of the bombs; therefore, the mass of the sample aliquots were uncertain for those samples, and spike recoveries could not be positively calculated. Based on the four samples that were not in question, recoveries ranged from 77 to 90 percent with an average of 85 and a standard deviation of 6. The percent recoveries for the sand substrate ranged from 89 to 104 percent, with an average of 97 and a standard deviation of 5.
Percent Recovery Clay Substrate | Percent Recovery Organic Substrate | Percent Recovery Sand Substrate | |
---|---|---|---|
118 | 99 | ||
91 | 93 | ||
132 | 97 | ||
144 | 98 | ||
79 | 89 | ||
176 | 104 | ||
102 | 88 | 95 | |
99 | 77 | 92 | |
111 | 84 | 98 | |
75 | 90 | 101 | |
Average | 113 | 85 | 97 |
Standard deviation | 31 | 6 | 5 |
This method is suitable for the determination of HgT in suspended solids isolated onto pre-baked quartz fiber filters (QFF). The minimum sample mass detectable is 0.059 ng and the range can be extended to meet most samples by controlling the amount of water filtered through the QFF and through dilutions of the oxidized sample. All samples are reported in ng/L.
Particulate samples collected onto baked quartz fiber filters are oxidized with BrCl to convert all forms of Hg to Hg2+. After oxidation, the sample is pre-reduced with NH2OH*HCl to remove the free halogens, then reduced with SnCl2 to convert Hg2+ to Hg0. The Hg0 is purged, captured on a gold trap, then thermally desorbed and detected using a CVAFS. This method can be used to determine HgT concentrations in particulate samples with a MDL of 0.059 ng of Hg on a filter. The amount of Hg on a filter is dependent on the solids load in the sample and volume of sample filtered.
To provide reliable concentrations for HgT in suspended solids, the WDML has developed a method to concentrate suspended solids from unfiltered water by either in-line filtration or by vacuum filtration. In either case, samples should be collected using clean hands/dirty hands sampling protocols (Olson and DeWild, 1999) to ensure sample integrity. Sample size is dependent on suspended particulate load and HgT concentration, but generally ranges from 0.5 to 1 liter. The suspended solids are retained on baked (prepared by heating to 550°C for four hours) QFF. To provide HgT concentrations for suspended solid samples in mass per unit volume (for example, ng/L), field crews must measure and record the volume of water filtered. Individual filters are placed into stackable Teflon petri dishes, double bagged in sealable plastic bags, and frozen. Samples are shipped to the lab on dry ice by overnight mail, and held frozen at a temperature of -15°C or less until processing. The WDML has not performed a holding time study; however, a frozen CRM certified for Hg is available through NIST and is stable for a duration of 9 years.
Samples are reported in ng/L. The ng/L concentration can be converted to ng/g if the mass of particles per volume of water is also determined. The ng/L concentration can also be combined with the ng/L concentration of the corresponding filtered water sample to obtain an estimate of the whole water Hg concentration. In hydrologic systems that are low in suspended solids, the filtered and whole water Hg concentrations can be indistinguishable within the limits of detection. The ability to separate and concentrate the suspended solids increases the ability to accurately evaluate the difference between the filtered and whole waters.
Suspended particulate loads vary considerably within hydrologic settings; therefore, varying amounts of sample water need to be filtered to obtain an appropriate amount of material for analysis. Using an MDL based on mass per filter allows a direct comparison of the amount of mercury in the oxidized sample solution to the consistent value of 0.059 ng rather than a constantly changing MDL expressed in ng/L based on the volume of sample water filtered. The volume of water filtered varies from 0.050 L to 1.0 L, and this range provides a minimum sample reporting limit varying from 0.059 ng/L to 1.18 ng/L.
To demonstrate that the WDML can accurately quantify HgT in particulate samples at or below environmentally important levels, an MDL study was performed. An MDL of 0.059 ng of HgT per filter was determined from multiple analyses of a CRM that was suspended in reagent water and filtered onto 10 separate QFFs (table 4). A mass of 0.0502 g of the CRM (MESS-2 marine sediment) was added to 2.0 L of reagent water. A subsample of the slurry was poured into a Hg clean Teflon bottle after thorough mixing. The remaining slurry was then mixed and filtered onto baked QFFs that were frozen until oxidation. Some of the filtrate was also collected into a clean Teflon bottle. The samples were analyzed over five days. The unfiltered and filtered water samples were analyzed during a single analysis run. For this study, the target volume per filter was 100.0 mL. This volume was not accurately achieved for each filter; therefore, the analytically determined mass per filter was converted to a 100.0 mL equivalency to accurately determine a mass-based MDL.
Table 4. Results from multiple analyses of particulate MESS-2 CRM for detection limit assay
[all masses in ng per 100.0 mL of slurry]
Particulate CRM | |
---|---|
0.201 | |
.208 | |
.177 | |
.184 | |
.164 | |
.152 | |
.141 | |
.169 | |
.176 | |
.187 | |
Average | .176 |
Standard deviation | .021 |
Percent relative standard deviation | 11.9 |
Detection limit (standard deviation x 2.821*) | .059 |
*students T-value at the 99 percent confidence interval for n=10
Upon receipt at the laboratory or on the day of preparation, reagent containers should be labeled with the date received or made and the initials of the person preparing them.
The stock and substock standards should be stored outside of the clean laboratory to prevent contamination of the entire lab. To clean a volumetric flask, fill to approximately 20 percent total volume with 50 percent HNO3, place the ground glass stopper on its side over the opening to prevent pressure buildup, and reflux on a hotplate for 4 hours.
The QCS will be prepared from a commercially available Hg standard. The standard will be of a different source or lot than the standard used to prepare the analytical standards. The QCS is used throughout analysis to verify statistical control.
Filters to be analyzed for HgT are thawed and placed into a 125-mL wide-mouth Hg clean Teflon bottles using a Teflon forceps. The Teflon petri dishes are rinsed three times with 5 percent BrCl into the sample bottle then the volume is brought up to 100.0 mL with the same 5 percent BrCl that was used for rinsing the petri dishes. The petri dish ID's are recorded and matched to the sample bottle ID that will now be used to track the sample through analysis. A method blank and a CRM are included with every ten samples, and a minimum of three blanks are included in each preparation batch. The preparation batch is all the samples that were prepared that day. A method blank includes a baked QFF and 100.0 mL of the 5 percent BrCl. A CRM is prepared by adding a measured mass of CRM to a 125 mL wide mouth Hg clean Teflon bottle and adding 100.0 mL of the 5 percent BrCl. The samples are capped tightly, double bagged and placed into an oven at 50°C for five days.
After oxidation, the samples are cooled, mixed, and 50 mL of sample is poured into a 60-mL Teflon vial containing 50 µL of NH2OH*HCl. The samples are analyzed according to USEPA Method 1631, Revision E (U.S. Environmental Protection Agency, 2002) utilizing an automated flow injection system. An analytical batch consists of 12 samples and is bracketed by the appropriate high level and low level QCS. All samples must fall within the calibration range. If they do not, they must be diluted appropriately. A sample can consist of a method blank, a CRM, or an environmental sample.
DQOs have been established to evaluate statistical control over the analytical system during the analysis.
where
RQCS = percent recovery for the QCSThe following formulas are used to determine the amount of Hg in ng/L from the original sample.
where
COx = concentration of oxidized solution, in ng/L
PA = peak area
S = slope of calibration line
where
MFB = mass of Hg in method blank, in ng
COx = concentration of oxidized solution, in ng/L, as determined in Equation 11
VB = volume of sample originally in bottle, in L
where
MF = mass of Hg on filter, in ng
COx = concentration of oxidized solution, in ng/L, as determined in Equation 11
VB = volume of sample originally in bottle, in L
MFB–AVE = average mass of Hg in all method blanks, in ng
where
CP = concentration of sample, in ng/L
MF = mass of Hg on filter, in ng
VF = volume of water that passed through filter, in L
Included with the preparation batch of environmental samples are method blanks and CRM samples. Each of these samples provides quality-control information used to evaluate the acceptability of the analytical runs. Typically, multiple analytical runs are necessary to analyze the entire preparation batch.
where
BDL = Batch detection limitTo evaluate method performance, two artificial water samples were created by adding different CRMs, 0.20814 g IAEA-405 polluted marine sediment and 0.0502 g MESS-2 marine sediment, to 2.0 L reagent water. IAEA-405 (certified concentration of 810 ng/g) represents environmental samples at higher concentrations, and MESS-2 (certified concentration of 92 ng/g) represents environmental samples at lower concentrations. Each of these artificial waters was filtered onto baked QFF filters to mimic environmental sample processing. Aliquots of the unfiltered samples were collected and analyzed to determine the beginning concentration of the solutions. Filtrate was also collected during sample processing and analyzed to determine how much mercury from the CRM dissolved in the reagent water.
Precision was evaluated by examining the percent relative standard deviation of the concentrations obtained from all analyses of the particulate filters for each concentration range. The percent relative standard deviation for the MESS-2 was 11.9 and for the IAEA-405 was 5.6 (tables 5 and 6). Percent relative standard deviations were calculated using the formula:
Table 5. Results for the analysis of MESS-2 CRM solution
Particulate Hg (ng/L) | Actual Percent Recovery | Theoretical Percent Recovery | |
---|---|---|---|
2.01 | 149 | 114 | |
2.08 | 155 | 118 | |
1.77 | 132 | 101 | |
1.84 | 137 | 105 | |
1.64 | 122 | 93 | |
1.52 | 113 | 87 | |
1.41 | 104 | 80 | |
1.69 | 126 | 96 | |
1.76 | 131 | 100 | |
1.87 | 139 | 106 | |
Average | 1.76 | 131 | 100 |
Standard deviation | 0.21 | 15 | 12 |
Percent relative standard deviation | 11.9 |
Table 6. Results for the analysis of IAEA-405 CRM solution
Particulate Hg (ng/L) | Actual Percent Recovery | Theoretical Percent Recovery | |
---|---|---|---|
76.3 | 103 | 95 | |
71.6 | 96 | 89 | |
70.3 | 94 | 87 | |
79.2 | 106 | 98 | |
75.4 | 101 | 93 | |
80.8 | 109 | 100 | |
82.9 | 111 | 103 | |
80.0 | 108 | 99 | |
80.3 | 108 | 99 | |
Average | 77.4 | 104 | 96 |
Standard deviation | 4.3 | 6 | 5 |
Percent relative standard deviation | 5.6 |
where
% RSD = percent relative standard deviation
CSTDEV = standard deviation of all samples
CAVE = average concentration of all samples
Accuracy was evaluated by calculating the theoretical and actual percent recovery of the CRM from each filter. The actual percent recovery utilized the analytically determined concentration for the unfiltered slurry, and the theoretical percent recovery used the mathematically derived concentration for the slurry based on the mass of the CRM added, the volume of water the CRM was added to, and the certified value for the CRM. Actual percent recoveries were calculated using the formula:
where
R = percent recovery
CP = analytically determined concentration of particulate sample
CUa = analytically determined concentration of unfiltered slurry
CF = analytically determined concentration of filtrate
Theoretical percent recoveries were calculated using the formula:
where
R = percent recovery
CP = analytically determined concentration of particulate sample
CUt = theoretical concentration of unfiltered slurry
CF = analytically determined concentration of filtrate
For the MESS-2 solution, the analytically determined concentration of the unfiltered slurry was 1.89 ng/L, the theoretical concentration of the unfiltered slurry was 2.30 ng/L, and the analytically determined filtrate concentration was 0.54 ng/L. Actual recoveries ranged from 104 to 155 percent, with an average of 131 and a standard deviation of 15 (table 5). The theoretical recoveries ranged from 80 to 118 percent, with an average of 101 and a standard deviation of 11. For the IAEA-405 solution, the analytically determined concentration of the unfiltered slurry was 78.0 ng/L, the theoretical concentration of the unfiltered slurry was 84.3 ng/L and the analytically determined concentration of the filtrate was 3.58 ng/L. Actual recoveries ranged from 94 to 111 percent, with an average of 104 and a standard deviation of 6 (table 6). Theoretical recoveries ranged from 87 to 103 percent, with an average of 96 and a standard deviation of 5.
This document describes the methods used by the Wisconsin District Mercury Lab to analyze solids and suspended solids for total mercury concentration. Because the procedures used to process and analyze solids and suspended solids differ, two distinct analytical performance studies were conducted and the results are presented.
The method detection limit established for the solids procedure as outlined in this report is 0.3 ng total mercury in a digestion bomb, which was acceptable because it is substantially below the levels commonly encountered in natural samples from a wide range of environments. The mass of sample digested varies, which gives a minimum sample reporting limit ranging from 0.6 ng/g to 6 ng/g. The method precision, calculated as the percent relative standard deviation, ranged from 10.0 to 12.8 percent. The accuracy of the procedure, which was determined from recovery tests for spiked samples, ranged from 85 to 113 percent and was acceptable.
The second method documented in this report is for sample preparation and analysis of suspended solids on baked quartz fiber filter. Two artificial raw water samples were created by suspending two different certified reference materials (IAEA-405 and MESS-2) in reagent water and filtering them to create suspended solids samples. The method detection limit was established by analyses of multiple filters of the MESS-2, and a limit of 0.059 ng was achieved. Based on the range of water filtered, the minimum sample reporting limit varies from 0.059 ng/L to 1.18 ng/L. Precision was evaluated by calculating percent relative standard deviation from analyses of replicate filters. Percent relative standard deviations were 5.6 and 11.9 for the IAEA-405 and MESS-2 certified reference materials, respectively. Accuracy was evaluated from percent recovery of total mercury on the filters from their target value. Target values were determined by subtracting the analytically determined concentration of the filtrate from the theoretical raw water sample concentration as well as the analytically determined raw water sample concentrations. Recoveries ranged from 94 to 155 percent, with an overall mean and standard deviation of 118 percent and 18, respectively, for the analytically determined concentrations. For the theoretical concentrations, the recoveries ranged from 80 to 118 percent, with an overall mean of 99 percent and a standard deviation of 9.
Biester, H., Gosar, M., and Covelli, S., 2000, Mercury speciation in sediment affected by dumped mining residues in the drainage area of the Idrija Mercury Mine, Slovenia: Environmental Science and Technology, v. 34, p. 3330–3336.
DeWild, J.F, Olson, M.L., and Olund, S.D., 2002, Determination of Methyl Mercury by Aqueous Phase Ethylation, Followed by Gas Chromatographic Separation with Cold Vapor Atomic Fluorescence Detection: U.S. Geological Survey Open-File Report 01–445, 14 p.
DeWild, J.F., Olund, S.D., Olson, M.L., and Tate, M.T., 2004, Methods for the Preparation and Analysis of Solids and Suspended Solids for Methylmercury: U.S. Geological Survey Techniques and Methods book 5, chap. A7, 13 p.
Gustin, M.S., Lindberg, S.E., Austin, K., Coolbaugh, M., Vette, A., and Zhang, H., 2000, Assessing the contribution of natural sources to regional atmospheric mercury budgets: The Science of the Total Environment, v. 259, p. 61–72.
Krabbenhoft, D.P., Wiener, J.G., Brumbaugh, W.G., Olson, M.L. DeWild, J.F., and Sabin T.J., 1999, A national pilot study of mercury contamination of aquatic ecosystems along multiple gradients, in Morganwalp, D.W., and Buxton, H.T., eds., U.S. Geological Survey Toxic Substances Hydrology Program—Proceedings of the technical meeting, Charleston, S.C., March 8–12, 1999—Volume 2 of 3—Contamination of hydrologic systems and related ecosystems: U.S. Geological Survey Water-Resources Investigations Report 99–4018–B, p. 147–160.
Mason, R.P., Fitzgerald, W.F., and Morel, F.M.M., 1994, The aquatic biogeochemical cycling of elemental mercury—Anthropogenic influences: Geochimica et Cosmochimica Acta, v. 58, p. 3191–3198.
Nater, E.A., and Grigal, D.F., 1992, Regional trends in mercury distribution across the Great Lakes states, north central USA: Nature, v. 358, p. 139–141.
Olson, M.L., and DeWild, J.F., 1999, Low-level techniques for the collection and species-specific analysis of low levels of mercury in water, sediment, and biota, in Morganwalp, D.W., and Buxton, H.T., eds., U.S. Geological Survey Toxic Substances Hydrology Program—Proceedings of the Technical Meeting, Charleston, South Carolina, March 8–12, 1999—Volume 2 of 3—Contamination of Hydrologic Systems and Related Ecosystems: U.S. Geological Survey Water-Resources Investigations Report 99–4018–B, p. 191–200.
U.S. Environmental Protection Agency, 1990, Guidelines establishing test procedures for the analysis of pollutants (Appendix B, Part 136, Definition of procedures for the determination of the method detection limit–Revision 1.11): Revised as of July, 1999, p. 537–539.
U.S. Environmental Protection Agency, 2002, Method 1631, Revision E: Mercury in water by oxidation, purge and trap, and cold vapor atomic fluorescence spectrometry: EPA-821-R-02-019, Office of Water, Washington D.C., 36 p.
Wiener, J.G., Krabbenhoft, D.P., Heinz, G.H., and Scheuhammer, A.M., 2003, Ecotoxicology of mercury, chapter 16 in Hoffman, D.J., Rattner, B.A., Burton, Jr., G.A., and Cairns, Jr., J., eds., Handbook of Ecotoxicology, (2d ed.): Boca Raton, Fla., CRC Press, p. 407–461.
Document Accessibility: Adobe Systems Incorporated has information about PDFs and the visually impaired. This information provides tools to help make PDF files accessible. These tools convert Adobe PDF documents into HTML or ASCII text, which then can be read by a number of common screen-reading programs that synthesize text as audible speech. In addition, an accessible version of Acrobat Reader 6.0, which contains support for screen readers, is available. These tools and the accessible reader may be obtained free from Adobe at Adobe Access.