Table 20. Detection limits, sensitivities, and calibration ranges for flame and graphite-furnace atomic absorption spectrometry

    Concentrationa, b in Solution for
Element Wavelength (nm) Detection Limit Sensitivityc Upper Limit of Linear Calibration Usable Nonlinear Range
Cda 228.8 0.05 0.065 5.0 5-20
Cub 324.7 0.02 0.07 5.0 5-15
Lib 670.8 0.02 0.035 2.0 2-10
Mgb 285.2 -- 0.008 0.5 0.5-2
Mnb 279.5 0.04 0.07 3.0 3-10
Nab 589.0 -- 0.02 1.0 1-5
Pbb 217.0 0.08 0.2 5.0 5-25
Znb 213.9 0.01 0.02 2.0 2-10
aGraphite-furnace AAS (ng/mL).
bFlame AAS (m g/mL).
cFor 0.0044 absorbance.