Link to USGS Home Page

The Chemical Analysis of Argonne Premium Coal Samples

Edited by Curtis A. Palmer
U.S. Geological Survey Bulletin 2144


Arsenic, antimony, and selenium by flow injection or continuous flow-hydride generation-atomic absorption spectrometry

By Philip L. Hageman and Eric Welsch

Table 1: Operating conditions for determination of As, Sb, and Se by HG-AAs

  Arsenic Antimony Selenium
Grating ultraviolet ultraviolet ultraviolet
Wavelength 193.7 to 197.2 nm 217.6 nm 196.0 nm
Slit 0.7 nm 0.2 nm 0.7 nm
Lamp setting 8W 8W 6W
Source EDL EDL EDL
Cell (furnace) temp 900°C 1,000°C 2,000°C
Mode absorbance absorbance absorbance
Strip chart recorder n/a n/a 10 mV, 5mm
Concentration mode, 10 mg 0.45A 0.20A 30A
next


[ Contents ]


Created by the EERT WWW Staff.

[an error occurred while processing this directive]