Determination of uranium concentration in water by liquid anion exchange-delayed neutron analysis
Links
- More information: Publisher Index Page (via DOI)
- Download citation as: RIS | Dublin Core
Abstract
Dissolved uranium is selectively removed from 11 of filtered, acidified water using a liquid anion exchange resin (Amberlite LA-1) dissolved in 10 ml of purified kerosene. The organic phase is then analyzed by a standard delayed neutron counting technique. Yields of removed uranium are consistently greater than 90 percent over a measured concentration range of 1.0 to 100 ppb uranium. The absolute detection limit based on 11 of water is 0.06 ppb. Elemental interferences are minimal and the results compare favorably with fluorometric analyses of natural waters.
Publication type | Article |
---|---|
Publication Subtype | Journal Article |
Title | Determination of uranium concentration in water by liquid anion exchange-delayed neutron analysis |
Series title | Journal of Radioanalytical and Nuclear Chemistry |
DOI | 10.1007/BF02036959 |
Volume | 84 |
Issue | 2 |
Year Published | 1984 |
Language | English |
Publisher | Springer Link |
Description | 6 p. |
First page | 207 |
Last page | 212 |
Google Analytic Metrics | Metrics page |