An automated solid-phase extraction (SPE) method was developed for the pre-concentration of chloroacetanilide and triazine herbicides, and two triazine metabolites from 100-ml water samples. Breakthrough experiments for the C18 SPE cartridge show that the two triazine metabolites are not fully retained and that increasing flow-rate decreases their retention. Standard curve r2 values of 0.998–1.000 for each compound were consistently obtained and a quantitation level of 0.05 μg/l was achieved for each compound tested. More than 10 000 surface and ground water samples have been analyzed by this method.
Suggested Citation
Meyer, M.T., Mills, M.S., Thurman, E., 1993, Automated solid-phase extraction of herbicides from water for gas chromatographic-mass spectrometric analysis: Journal of Chromatography A, v. 629, no. 1, p. 55-59, https://doi.org/10.1016/0021-9673(93)80353-A.
| Publication type | Article |
|---|---|
| Publication Subtype | Journal Article |
| Title | Automated solid-phase extraction of herbicides from water for gas chromatographic-mass spectrometric analysis |
| Series title | Journal of Chromatography A |
| DOI | 10.1016/0021-9673(93)80353-A |
| Volume | 629 |
| Issue | 1 |
| Year Published | 1993 |
| Language | English |
| Publisher | Elsevier |
| Contributing office(s) | Toxic Substances Hydrology Program |
| Description | 5 p. |
| First page | 55 |
| Last page | 59 |