HIF evaluation of In-Situ Aqua TROLL 400
Links
- Document: Report (856 KB pdf)
- Download citation as: RIS | Dublin Core
Abstract
The In-Situ Aqua TROLL 400 (Aqua TROLL 400) was tested at the U.S. Geological Survey (USGS) Hydrologic Instrumentation Facility (HIF) against known standards over the Aqua TROLL 400’s operating temperature to verify the manufacturer’s stated accuracy specifications and the USGS recommendations for pH, dissolved oxygen (DO), and specific conductance (SC). The Aqua TROLL 400 manufacturer’s specifications are within the USGS recommendations for all parameters tested, except for DO, which is outside the USGS recommendation at DO concentrations of 8.0 milligrams per liter (mg/L) and higher. The Aqua TROLL 400 was compliant with Serial Digital Interface at 1200 baud (SDI-12) version 1.3. During laboratory testing of pH, the Aqua TROLL 400 sonde met the U.S. Geological Survey “National Field Manual for the Collection of Water-Quality Data” (NFM) recommendations for pH at all values tested, except at 4 degrees Celsius (°C) at pH 9.395 and pH 3.998. The Aqua TROLL 400 met the manufacturer specifications for pH at all values tested, except for pH buffers 3.998, 9.395, and 10.245 at 4 °C; pH 2.990 and 3.998 at 15 °C; and pH 3.040 at 40 °C. The Aqua TROLL 400 met the NFM recommendations at 93.7 percent of the SC values tested and met the manufacturer’s accuracy specifications at 56.3 percent of the SC values tested. During the laboratory testing for DO, the Aqua TROLL 400 met the manufacturer specifications, except at 5.55 mg/L, and met the NFM recommendations at all concentrations tested. An Aqua TROLL 400 was field tested at USGS Station 02492620, National Space Technology Laboratories (NSTL) Station, Mississippi, on the Pearl River for 6 weeks and showed good agreement with the well-maintained site sonde data for pH, DO, temperature, and SC.
Suggested Citation
Tillman, E.F., 2017, HIF evaluation of In-Situ Aqua TROLL 400: U.S. Geological Survey Open-File Report, 2017–1086, 35 p., https://doi.org/10.3133/ofr20171086.
ISSN: 2331-1258 (online)
Table of Contents
- Abstract
- Introduction
- Description
- Test Procedures
- Field Test
- Summary
- References Cited
Publication type | Report |
---|---|
Publication Subtype | USGS Numbered Series |
Title | HIF evaluation of In-Situ Aqua TROLL 400 |
Series title | Open-File Report |
Series number | 2017-1086 |
DOI | 10.3133/ofr20171086 |
Year Published | 2017 |
Language | English |
Publisher | U.S. Geological Survey |
Publisher location | Reston, VA |
Contributing office(s) | Office of Surface Water |
Description | vi, 35 p. |
Online Only (Y/N) | Y |
Additional Online Files (Y/N) | N |
Google Analytic Metrics | Metrics page |