COULOMETER SET-UP


PROCEDURE FOR STARTING THE COULOMETER
Materials Required: Preparation materials for starting Coulometer; link to larger image

Fill titration cell:
Cathode Solution goes in large section.
Anode Solution goes in small section.

Add KI Granular to Anode Solution until it covers the bottom of the section - insert stir bar into large section.

Titration cell with chemicals; link to larger image

Fill KOH, KI, and AgNO3 traps 1/2 full.
The KOH trap is the glass vial with the red cap.
The KI trap is the glass vial with the yellow tint.
The AgNO3 trap is the glass vial with the grayish tint.

Add 2-3 drops of 2N H2SO4 to the KI and AgNO3 traps.

Add drops of acid to traps; link to larger image

Attach the nitrogen gas tube to the KOH trap.

Connect gas tube to KOH trap; link to larger image

Turn on nitrogen gas (large silver knob) and increase gas flow (approximately 5 psi) using the blue Increase/Decrease knob on the gas tank regulator.

Set flow control at 100 cc/min on the flow meter on the heating unit.

Adjust gas flow; link to larger image

Connect KOH trap to reaction vial.


KOH traps CO2 out of the gas.

Connect KOH trap to reaction vial; link to larger image

Connect the reaction vial to the thin section of the AgNO3 trap.


AgNO3 traps out H2S.

Reaction vial connected to AgNO3 trap; link to larger image

Connect the top of the AgNO3 trap to the thin section of the KI trap.


KI traps out weak acids.

Silver Nitrate trap connected to KI trap; link to larger image

Connect the top of the KI trap to the titration cell.

This is what the completed set-up should look like.

Coulometer set up; link to larger image

Teflon tape all joints to avoid leaks.

Teflon tape all joints; link to larger image

Turn on the main power to the heating unit.

Turn on heating unit; link to larger image

Turn on the main power to the coulometer.

Main power switch; link to larger image

Turn on the water bath - set at 11 degrees C.

Turn on water bath; link to larger image


Make sure to open the return valve before opening the source valve.

Turn on return valve first; link to larger image

Rotate the titration cell until maximum transmittance (%T) is attained.

Check the system for bubble transmittance to the titration cell.

Rotate the titration cell; link to larger image

Adjust transmittance to 100%.

Never touch the titration cell after this point.

Adjust transmittance; link to larger image

Remove jumper cable and connect titration cell to electrodes:
Red to red
Black to black.

Connect electrodes to titration cell; link to larger image

Turn the cell current on and allow the cell current to titrate the cell solution to its endpoint.

The solution color will turn blue when the %T reaches 29.

 

Titration cell turned blue; link to larger image

Reset the coulometer using Reset Button.

Reset button on Coulometer; link to larger image

Set the timer for 10 minutes.

Timer set for 10 minutes; link to larger image