USGS - science for a changing world

Background Data Acquired from Gamma-ray Castle

  


 

Background Measurements

 

The background count rates for the gamma-ray spectrometer are measured at intervals of 15 to 30 days.  Normally, the background measurements are made multiple times over a weekend using counting times of 8 hours.  Figure 1 shows the count rate (counts per second) for the reference source of Cs137 that is used to stabilize the spectrum.  These data indicate that the system has a slow drift over time with the count rates slowly increasing.  The change that occurs in August, 2001 corresponds to an adjustment made to the energy calibration of the spectrometer.

 

Cesium-137 background counts versus time

 Figure 1:  Cs137 background count rates vs. Time

 

Figure 2 shows the background count rates for the K40 channel.  The data do not demonstrate any clear long term trends.  A close examination of the consecutive measurements does suggest that the background count rates vary linearly over periods of a few days.  The largest variations are about 130 counts per hour.  A count rate of 130 counts per hour corresponds to about 0.03 percent K for a 600 gram sample.

Potassium-40 background counts versus time

 

Figure 2:  K40 background count rates vs. Time

 

Figure 3 show the background count rates for the Bi214 (uranium decay series) channel.  These data are very similar to the K40 data with maximum variations on the order of 80 counts per hour over a period of a few days.  A count rate of 80 counts per hour corresponds to about 0.2 ppm eU for a 600 gram sample.

Bismuth-214 background counts versus time

 

Figure 3:  Bi214 background count rates vs. Time

 

Figure 4 shows the count rates for the Tl208 (thorium decay series) channel.  These data are also very similar to the K40 data with maximum variations on the order of 80 counts per hour over a period of a few days.    A count rate of 80 counts per hour corresponds to about 0.7 ppm eTh for a 600 gram sample.

Thallium-208 background counts versus time

 

Figure 4:  Tl208 background count rates vs. Time


Data for Some USGS Rock Standards Acquired from Gamma-ray Castle

The figures below show the calibration equations in graphical terms as a means of comparing the calculated values versus the "known" values for some USGS rock standards.  CKn, CUn, and CThn are the count rates after background correction and normalized to counts per hour per 600 grams of sample material.

 

The equations are:

Parameters After Fitting:

A =  0.000223

B = -0.000157

C = -0.000072

 

Calculated Potassium Values vs. Known
Potassium Values

 

Figure 5:  Calculated Potassium Values vs. Known Potassium Values

(Click on image to view larger image in new window)

 

Parameters After Fitting:

A = -0.0000083

B =  0.00252

C = -0.00062

 

Calculated Uranium Values vs. Known
Uranium Values

 

Figure 6:  Calculated Uranium Values vs. Known Uranium Values

(Click on image to view larger image in new window)

 

Parameters After Fitting:

A = -0.000003

B = -0.000164

C =  0.00854

 

Calculated Thorium Values vs. Known
Thorium Values

 

Figure 7:  Calculated Thorium Values vs. Known Thorium Values

(Click on image to view larger image in new window)



Top  ||  Introduction  ||  Software  ||  Hardware  ||  Construction  ||  References  ||  Photos  ||  Data 


For more information about this report contact: Stephen L. Snyder or Joseph S. Duval
[an error occurred while processing this directive]