USGS Open-File Report 2004-1443, Operation Manual: Time-Series, Storm-Activated Suspended Sediment Sampler Deployed in the Coastal Ocean
Back to Main Appendix
Appendix 1
McLane
Research Laboratories, USA
Water Transfer System ‑ 18
Program PUMP V1.42
**************************
* MAIN MENU *
**************************
S/N 521 ‑‑ transmissometer version
<1> Set clock
<2> Display time,
temperature & voltage
<3> Run pump motor
<4> Move valve
<5> Preload filters
<6> Create schedule
<7> Run schedule and
deploy
<8> Offload data to a
disk
<9> Display data from a
disk
<X> Exit program
<ESC> Call MAIN MENU
Computer Font Select a number | User
Font 7
|
Initial pump flow rate (mL/minute) ‑ ‑ ‑ ‑ ‑
‑ ‑ ‑ ‑ (50‑125) | 100 |
Total flow (mL) ‑ ‑ ‑
‑ ‑ ‑ ‑ ‑ ‑ (10‑100000) | 3000 |
Minimum pump flow rate (mL/minute) ‑ ‑ ‑ ‑ ‑
‑ ‑ ‑ ‑ (20‑100) | 50 |
Maximum pumping time (minutes) ‑ ‑ ‑ ‑
‑ ‑ ‑ ‑ ‑ (1‑7200) | 120 |
Begin reading transmissometer
data YEAR ‑ ‑ ‑ ‑
‑ ‑ (0‑99) | 1 |
Begin reading transmissometer
data MONTH ‑ ‑ ‑ ‑ ‑ (1‑12) | 10 |
Begin reading transmissometer
data DAY ‑ ‑ ‑ ‑ ‑ ‑ (1‑31) | 24 |
Begin reading transmissometer
data HOUR ‑ ‑ ‑ ‑
‑ ‑ (0‑23) | 8 |
Stop reading transmissometer data YEAR ‑ ‑ ‑ ‑ ‑
‑ (0‑99) | 2 |
Stop reading transmissometer data MONTH
‑ ‑ ‑ ‑ ‑ (1‑12) | 2 |
Stop reading transmissometer data DAY
‑ ‑ ‑ ‑ ‑ ‑ (1‑31) | 4 |
Stop reading transmissometer data HOUR ‑ ‑ ‑ ‑ ‑
‑ (0‑23) | 8 |
Use transmissometer to
trigger pump (1=yes 0=no) ‑ ‑
(0‑1) | 1 |
Minimum transmissometer trigger
threshold (mV) ‑ ‑ ‑
(0‑5000) | 3500 |
Number of summed
transmissometer intervals ‑ ‑
‑ ‑ ‑ (2‑960) |
128 |
First to middle storm
event maximum time (hrs) ‑ ‑ (1‑1000) | 72 |
Middle to final storm
event maximum time (hrs) ‑ ‑ (1‑1000) | 136 |
Max number of storms to
sample ‑ ‑ ‑ ‑ ‑
‑ ‑ ‑ ‑ ‑ ‑ (0‑6) | 4
|
Computer Font | User Font
|
50% deployment no storm
threshold reduction (%) ‑ ‑ ‑ (0‑100) | 10 |
75% deployment no storm
threshold reduction (%) ‑ ‑ ‑ (0‑100) | 5 |
85% deployment no storm
threshold reduction (%) ‑ ‑ ‑ (0‑100) | 3 |
Time between ON and OFF dates = 103 day(s) 0 hour(s)
A/D data will be stored every 1540 seconds [ 25 min(s) 40 sec(s) ]
Please standby ‑ sending information to instrument ....
...................................Verified OK
|