USGS Open-File Report 2004-1443, Operation Manual: Time-Series, Storm-Activated Suspended Sediment Sampler Deployed in the Coastal Ocean
Click here for Table 1 in Excel format.
Table 1. Results of suspended matter concentrations (mg/l) collected using filter holders from the WTS-Tr instrument with the internal top frit in place and removed. Results from the standard laboratory millipore filtration system were used as a control.
The coefficient of variation is the standard deviation expressed as a percentage of the mean.
Frit
Configuration |
Weight of suspended matter (mg) |
Volume filtered ( ml) |
mg/l |
Mean |
Standard deviation |
Coefficient of variation (%) |
Frit in Place |
0.829 |
493 |
1.68 |
|
|
|
|
0.499 |
413 |
1.21 |
1.37 |
0.27 |
19.59 |
|
0.640 |
525 |
1.22 |
| | |
|
Frit Removed |
1.11 |
493 |
2.26 |
| | |
|
1.22 |
495 |
2.46 |
2.44 |
0.17 |
7.00 |
|
1.37 |
525 |
2.60 |
| | |
|
Millipore |
1.42 |
535 |
2.65 |
| | |
Rig |
1.45 |
520 |
2.79 |
2.83 |
0.21 |
7.36 |
|
1.59 |
520 |
3.06 |
| | |
|
|